Sign In | Join Free | My disqueenfrance.com
China China World Technology Medical Equipment Service Group logo
China World Technology Medical Equipment Service Group
China World Technology Medical Equipment Service Group Win-win together!
Verified Supplier

5 Years

Home > Defibrillator Machine Parts >

Cardiolife TEC-7621C TEC-7721C Defibrillation Printer Recorder Unit WS-761V

China World Technology Medical Equipment Service Group
Trust Seal
Verified Supplier
Credit Check
Supplier Assessment
Contact Now

Cardiolife TEC-7621C TEC-7721C Defibrillation Printer Recorder Unit WS-761V

Brand Name : Nihon Kohden

Model Number : TEC-7621C TEC-7721C

MOQ : 1

Payment Terms : T/T, Western Union,PayPal

Delivery Time : 3-5 working days

Product name : Nihon Kohden Defibrillator

Type : TEC-7621C TEC-7721C

Application : medical equipment,ICU,hospital device

Warranty : 90days

Mode of transport : Express delivery

Nominal Voltage : 110V—240v

Frequency : 50/60 Hz ~

Supported language : English,French,Hindi,Portuguese,Russian,Arabic,etc

Service : Professional technical team

Apply For : Hospital, Clinic, Hospital room, ICU, Sickroom, Ward, Bed-ward, Inpatient ward

waveform recording : 50, 25, 5mm/s

Contact Now

NIHON KOHDEN Cardiolife TEC-7621C TEC-7721C defibrillation printer Recorder Unit WS-761V

► Brief Introduction

This printer Recorder Unit WS-761V is 3-channel built in thermal array recorder records various defibrillation information. And work for NIHON KOHDEN Cardiolife TEC-7621C TEC-7721C defibrillation.

Remove the recorder unit

  1. Remove the battery pack, memory card and lower case.
  2. Remove the main chassis and main board.
  3. Remove the two M3 pans, tighten the screws with spring washers, and pull out the recorder unit.

Cardiolife TEC-7621C TEC-7721C Defibrillation Printer Recorder Unit WS-761V

► Record Type

  • Manual recording: real-time/delayed waveform recording, report recording, event recording
  • Automatic record: charge record after discharge, alarm record, regular record
  • Paper speed: real-time/delayed ECG waveform recording: 50, 25, 5mm/s

► Check Recorder Screen
You can check the recorder. When “4.Check Recorder” on the System Maintenance - Check Hardware screen in selected and the OK Button is pressed, the recorder prints test patterns.
After all test patterns are printed, the System Maintenance - Check Hardware screen is displayed.

Cardiolife TEC-7621C TEC-7721C Defibrillation Printer Recorder Unit WS-761V

Troubleshooting

Problem Possible Cause Action
Printing is blurred. Dots are missing The specified recording papar is not used Use the specified recording paper.
The thermal head is dirty. Clean the thermal head with the head cleaning pen. If the troble not solved. The recorder unit is faulty. Replace the recorder unit.
There is no printing The recording paper is not loaded. Load new recording paper.
Recording unit door is not properly closed. Close the door until it clicks.
The recording paper is set with the wrong side facing up. Set the recording paper correctly.
Faulty key board. If the lamp dose not light when you press the record key, cleck the key function and recorder with the System Maintenance screen. If the recorder check is passed, the key board is faulty. Replace the key board.
Faulty recorder If the lamp lights when you press the record key, check the recorder in the System Maintenance screen.
Faulty main board
Printing is light. The recording unit temperature is too hot. Cool the instrument down in cooler place. If the trouble is not solved, the recorder unit or main board is faulty. Replace the recorder unit or main board.

► Delivery Issues

As a general rule, all the orders are going to be shipped within 3 to 5 working days and delivered on time. Of course sometimes there are some special circumstances. For instance, on busy days the shipment will be postponed. But no worries, all the orders will be dispatched finally. We will make great efforts to process the orders and ship them out as soon as possible. Please stay patient and keep in touch with us. If you need the product to be shipped urgently, please make sure you talk to us in advance before placing your orders. In order to make the deal runs successfully, please make it a point and manage your time of placing orders on the basis of your business needs. If you have any questions such as your order is taking much more time than expected or you need to know when will your order arrive, you can contact us at anytime. We will reply you as soon as possible. Your satisfaction is our biggest wish.

If you are interested in Hospital Medical Equipment Used NIHON KOHDEN Cardiolife TEC-7621C TEC-7721C defibrillation printer Recorder Unit WS-761V , welcome to place your orders.


Product Tags:

TEC-7721C Defibrillation Printer Recorder Unit

      

Defibrillation Printer Recorder Unit WS-761V

      

TEC-7621C Defibrillation Printer Recorder Unit

      
China Cardiolife TEC-7621C TEC-7721C Defibrillation Printer Recorder Unit WS-761V wholesale

Cardiolife TEC-7621C TEC-7721C Defibrillation Printer Recorder Unit WS-761V Images

Inquiry Cart 0
Send your message to this supplier
 
*From:
*To: China World Technology Medical Equipment Service Group
*Subject:
*Message:
Characters Remaining: (0/3000)
 
Inquiry Cart 0